34 GOSSEN METRAWATT GMBH
11.7 Testing Extension Cables for VDE 0701, Part 1, and VDE 0702
(option: EL1 adapter)
Extension Cables up to 5 m Long
Protective conductor resistance between the earthing contact at the
mains plug and all exposed metal parts may not exceed 0.3 Ω for safety
class I devices. This value may not exceed 1 Ω for permanently
connected data processing systems (DIN VDE 0701, part 240).
Extension and Connector Cables Longer than 5 m
Per DIN VDE 0702, an additional cable resistance of 0.1 Ω, however, not
more than 1 Ω, is allowable as of a length of 5 m for each additional
7.5 m.
Resistance testing for cables longer than 5 m is thus advisable (see also
limit values on page 17).
Note
The EL1 accessory adapter is required for testing for short-
circuiting and interruption of single-phase extension cables.
Connecting the Extension Cable or the Multiple Outlet
Note
The keys in the adapter’s handle have no function.
Performing the Test
Connect the extension cable to the EL1 adapter as shown in the figure
above.
Select the “Ext. Cable” test from the initial menu with the cursor and
acknowledge with the
key: Ext. Cable
x
WITH EL1.
Select “Start Test” with the
key.
Start the measurement sequence with the
key.
First perform and acknowledge visual inspection of the extension cable.
Enter the length of the cable with the
and
keys. Acknowledge with
the
key.
PROBE 5 4
(21)
EL1
SECUTEST
Device Under Test =
Extension Cable